MPT-AES的注射泵-超声雾化器联用进样系统

冯国栋, 韩松柏, 于爱民   

  1. 吉林大学化学学院, 长春 130012
  • 收稿日期:2005-01-26 修回日期:1900-01-01 出版日期:2005-07-26
  • 通讯作者: 于爱民

Studies on Injector Pump-ultrasonic Nebulizer Feeding Sample System for MPT-AES

FENG Guo-dong, HAN Song-bai, YU Ai-min   

  1. College of Chemistry, Jilin University, Changchun 130012, China
  • Received:2005-01-26 Revised:1900-01-01 Online:2005-07-26
  • Contact: YU Ai-min

摘要: 报道了用于微波等离子体炬原子发射光谱仪(MPT-AES)的新型注射泵-超声 雾化器联用进样系统, 考察了影响该系统性能的因素. 分别用气动雾化进样系统和注射泵- 超声雾化联用进样系统进样, 测定了Cd, Cu, Mg, Ca和Na等5种元素、 浓度均为1×10 -4%的混合溶液. 结果表明, 注射泵-超声雾化联用进样系统所得的谱线强度是 气动进样所得谱线强度的2~3倍, 而超声雾化进样的检出限比气动雾化进样小2~10倍 , 且超声雾化进样系统的记忆效应较强, 所需清洗时间较长.

关键词: 注射泵-超声雾化联用进样, MPT-AES, 气动雾化进样

Abstract: The feeding-sample system of MPT-AES is pneumatic nebulization syste m, but its efficiency is not good. A new-built injector pump-ultrasonic nebuli zer combination feeding-sample system has been designed. Its performance was te sted and compared with that of the pneumatic nebulization system. It can be conc luded that the new-built feeding-sample system can increase the spectral line intensity about 2~3 times and decrease the detection limit about 2~10 times. M oreover, this new-built system can reduce the time taken by washing the sample c ell from 30 min or so to about 10 min.

Key words: injector pump-ultrasonic nebulizer, MPT-AES, pneumatic nebulization system

中图分类号: 

  • O652.2