J4

• 化学 • 上一篇    下一篇

偏最小二乘法用于近红外漫反射光谱测定异福片

逯家辉, 吴凌, 郭伟良, 田晓乐, 陈渝飞, 石健, 滕利荣   

  1. 吉林大学 生命科学学院, 长春 130012
  • 收稿日期:2006-08-31 修回日期:1900-01-01 出版日期:2007-05-26 发布日期:2007-05-26
  • 通讯作者: 滕利荣

Rapid Determination of Rifampicin and Isoniazide Tablets UsingNear Infrared Diffuse Reflectance Spectroscopy Combined with Partial Least Square

LU Jiahui, WU Ling,  GUO Weiliang, TIAN Xiaole, CHEN Yufei, SHI Jian, TENG Lirong   

  1. College of Life Science, Jilin University, Changchun 130012, China
  • Received:2006-08-31 Revised:1900-01-01 Online:2007-05-26 Published:2007-05-26
  • Contact: TENG Lirong

摘要: 应用近红外漫反射光谱结合偏最小二乘法(NIRPLS)对异福片中利福平和异烟肼的含量进行定量分析, 所建立的定量分析模型对校正集样品中利福平和异烟肼的含量预测回归系数分别为0.992 77,0.989 01,交互验证均方根误差(RMSECV)分别 为0.006 65,0.004 23.对预测集样品利福平、 异烟肼的含量预测均方根误差(RMSEP)分别为0.005 73,0.003 79;加样平均回收率分别为99.376%和98.243%;重现性实验相对标准偏差(RSD)分别为0.679 3%和0.639 8%.结果表明, 该方法预测精度高, 且具有方便快捷、 非破坏、 无污染、 可在线检测、 重现性好等优点, 可作为异福片原位质量检测和在线质量监控的方法予 以推广.

关键词: 近红外漫反射光谱, 偏最小二乘法, 利福平, 异烟肼

Abstract: Near infrared diffuse reflectance spectroscopy combined with partial least square (NIRSPLS) was applied to establishing calibration models for determining the contents of rifampicin and isoniazide in rifampicin and isoniazide tablets.The regression coefficient (R) by means of the model for determining the content of rifampicinis 0.992 77, the root mean square error of crosscalibration (RMSECV) is 0.006 65, the R by means of the model for determining the content of isoniazide is 0.989 01,  and the RMSECV is 0.00423.Using these models for predicting the contents of rifampicin and iso niazide in validation set, the root mean square error of prediction (RMSEP) are 0.005 73 and 0.003 79.These results demonstrate that this method possesses high precision of prediction. The average recoveries are 99.376% and 98.243%in the recovery experiments, and the relative standard deviation (RSD) are 0.679 3% and 0.639 8% in reappearance experiments. With its advantages of facility, rapid, nondestruction, no pollution, online detection and good reappearance, this method must be popularized in situ measurement and in the online quality control for rifampicin and isoniazide tablets production.

Key words: near infrared diffuse reflectance spectroscopy, partial least square, rifam picin, isoniazide

中图分类号: 

  • O657.33