J4
• 化学 • Previous Articles Next Articles
LU Jia-hui, JIANG Chao-jun, MENG Qing-fan, GAO Chao-hui, WEI Wei, ZHAO Hang, TENG Li-rong
Received:
Revised:
Online:
Published:
Contact:
Abstract: We have applied partial least squares (PLS) to establish a calibration model, the precision of which is verified by cross-validation. Via adopting optimum wavelength to calibrate, a satisfactory model has been established. In this model, the correlation coefficient in methanol determination is 0.999 91, the root mean square error of cross-validation (RMSECV) is 0.431; in ethanol de termination, the correlation coefficient is up to 0.999 98, and the RMSECV is 0.193. Compared with that of gas chromatography, the relative error of the sample determination results is not more than 3.505%.
Key words: near infrared spectrum, partial least squares, methanol, ethanol
CLC Number:
LU Jia-hui, JIANG Chao-jun, MENG Qing-fan, GAO Chao-hui, WEI Wei, ZHAO Hang, TENG Li-rong. Determination of Methanol and Ethanol in Ternary Mixture by Near Infrared Spectroscopy with PLS Regression[J].J4, 2005, 43(03): 368-371.
0 / / Recommend
Add to citation manager EndNote|Reference Manager|ProCite|BibTeX|RefWorks
URL: http://xuebao.jlu.edu.cn/lxb/EN/
http://xuebao.jlu.edu.cn/lxb/EN/Y2005/V43/I03/368
Cited