MCNP Simulation of High-Purity Germanium Detector Efficiency

ZHANG Jian fang1, ZHAO Guang yi1, WANG Yu de1, WANG Ming yong2, MA Yu gang1, ZHANG Hui1   

  1. 1. College of Physics, Jilin University, Changchun 130012, China;2. Department of Physics, Anshan Normal College, Anshan 114005, Liaoning Province, China
  • Received:2009-11-16 Online:2010-09-26 Published:2010-09-21
  • Contact: ZHAO Guang yi E-mail:zhaogy@jlu.edu.cn

Abstract:

The experimental detection efficiency of 152Eu and 133Ba at 15 cm was measured via coaxial high-purity germanium (HPGe) detector.  Then the coaxial HPGe detector peak efficiency was simulated via Monte Carlo method, to adjust detector dead layer thickness and cold\|finger size. The results show that when the HPGe detector dead layer thickness is 0.22 cm, the thickness and length of cold\|finger size are 0.301 cm and 1 cm, the simulated efficiency is consistent with the experimental efficiency.

Key words: high-purity germanium detector, peak detection efficiency, Monte Carlo method, dead layer thickness, cold-finger size