吉林大学学报(工学版) ›› 2017, Vol. 47 ›› Issue (4): 1159-1164.doi: 10.13229/j.cnki.jdxbgxb201704021

• Orginal Article • Previous Articles     Next Articles

Standard generator and its calibration method for analog component video signal with distortion

CHEN He1, YANG Zhi-jun2, WU Zhao-chun2, Feng Zhi-gang2, ZHAO Ke-jia2, LI Di-fei2, TIAN Di1   

  1. 1.College of Instrumentation & Electrical Engineering, Jilin University, Changchun 130061, China;
    2.Division of Electronics and Information Technology, National Institute of Metrology, China, Beijing 100029, China
  • Received:2016-06-23 Online:2017-07-20 Published:2017-07-20

Abstract: When standard analog component video test signal is employed to calibrate video signal analysis instrument, the measurement results are insufficient. To solve this problem, first, the measurement principle of analog component video signal was analyzed. Then, based on the SDP2000 signal editing software and TG700 video signal generator, analog component video test signal with distortion was developed. A standard video signal source was set up, whose voltage level and rise time were calibrated and the measurement uncertainty was analyzed. Results show that the expended uncertainty of voltage level is 0.60% and the expended uncertainty of rise time is 0.19% (coverage factor k=2). Such standard source is able to generate standard definition and high definition analog component video test signal with distortion, which can be used to calibrate video signal analysis instrument to obtain various results and evaluate the measurement accuracy of the instrument more comprehensively.

Key words: instrument and meter technology, distortion analog component video test signals, digital signal synthesis, instrument calibration, uncertainty of measurement

CLC Number: 

  • TN98
[1] 陈善栘. 数字视频测量应用技术(基础篇)[M]. 北京: 人民邮电出版社, 2012: 88-92.
[2] 孙永生, 胡菊萍, 钟伟军. SDI视频信号分析仪格式监视功能校准研究[J]. 广播电视监测, 2008, 35(1): 99-103.
Sun Yong-sheng, Hu Ju-ping, Zhong Wei-jun. Study of calibration for format monitoring function of SDI vide signal analyzer[J].Radio and TV Monitoring, 2008, 35(1): 99-103.
[3] 孙永生. 数字电视分量视频信号测试仪器校准技术研究[D]. 桂林: 桂林电子科技大学电子工程与自动化学院, 2008.
Sun Yong-sheng. Study on calibration technology for component video signal test instruments of DTV[D]. Guilin: College of Electrical Engineering and Automation, Guilin University of Electronic Technology, 2008.
[4] 杨智君, 吴昭春, 马晓庆, 等. 视频失真校准方法的研究[J]. 计量学报, 2012, 33(2): 158-162.
Yang Zhi-jun, Wu Zhao-chun, Ma Xiao-qing, et al. The development of the video distortion calibration[J]. Acta Metrologica Sinica, 2012, 33(2): 158-162.
[5] 刘蕾, 韩东. VM700T视频测量仪的校准方法的改进[J].电视技术, 2012, 36(16): 94-97.
Liu Lei, Han Dong. Improved calibration method of VM700T video measurement set[J]. Video Engineering, 2012, 36(16): 94-97.
[6] 杨智君, 马晓庆, 吴昭春. 模拟/数字视频校准技术培训教材[M]. 北京: 中国计量科学研究院,信息与电子计量科学与测量技术研究所, 2012: 49-66.
[7] Tektronix. VM series video measurement system user manual[EB/OL].[2016-04-26].http://cn.tek.com/vm5000hd-manual/vm -series.
[8] 倪育才. 实用测量不确定度评定[M]. 4版. 北京:中国质检出版社, 2014:78-112.
[9] ISO/IEC GUIDE 98-3:2008(E). Uncertainty of measurement-part 3: Guide to the expression of uncertainty in measurement[S].
[10] LeCroy Corporation. LeCroy Waverunner 6 Zi oscilloscope datasheet[EB/OL].[2016-04-26]. http://www.metronic.dk/CustomerData/Files/Folders/6-lecroy-pdf/92_waverunner-6zi.pdf.
[11] NTC Tech.Tektronix TG700 TV signal generator platform user manual[EB/OL].[2016-04-26].http://www.zeus.go.kr/cafe/public/post/24304/down/1658.
[1] Han Cheng-hao,Zhang Zhong,Jiang Bao-lin,Li Min . Waste water activity determinator using constant wavelength sensing technology [J]. 吉林大学学报(工学版), 2007, 37(03): 701-0704.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
No Suggested Reading articles found!