Studies on Injector Pump-ultrasonic Nebulizer Feeding Sample System for MPT-AES

FENG Guo-dong, HAN Song-bai, YU Ai-min   

  1. College of Chemistry, Jilin University, Changchun 130012, China
  • Received:2005-01-26 Revised:1900-01-01 Online:2005-07-26
  • Contact: YU Ai-min

Abstract: The feeding-sample system of MPT-AES is pneumatic nebulization syste m, but its efficiency is not good. A new-built injector pump-ultrasonic nebuli zer combination feeding-sample system has been designed. Its performance was te sted and compared with that of the pneumatic nebulization system. It can be conc luded that the new-built feeding-sample system can increase the spectral line intensity about 2~3 times and decrease the detection limit about 2~10 times. M oreover, this new-built system can reduce the time taken by washing the sample c ell from 30 min or so to about 10 min.

Key words: injector pump-ultrasonic nebulizer, MPT-AES, pneumatic nebulization system