Control of Solasedine Hydrochloride Quality

LI Liying1, FAN Junheng2, LIU Liang1   

  1. 1. College of Chemistry and Biology, Beihua University, Jilin 132013, Jilin Province, China;2. Jilin University Press, Changchun 130021, China
  • Received:2008-12-12 Revised:1900-01-01 Online:2009-05-26 Published:2009-06-23
  • Contact: LI Liying

Abstract: Highperfomance liquid chromatography was used to study how to control solasedine hydrochloride quality. Column was Agilent ZORBAX Eclipse XDB-C18 column (150 mm×4.6 mm, 5 μm), mobile phase was Vmethanol∶Vacetonitrile∶V0.020 mol·L-1lammonium acetate=60 ∶35 ∶5, flow rate was 1 mL/min, detection wavelength 205 nm, injection volume 10 μL, the solasodine hydrochloride quality separated from other impurities in a good linear range of 5~500 μg/mL (r=0.999 3), numder of theoretical plates n>12 000(R>1.5); the average recovery for the sample 98.6% (n=9). Relative standard deviation of reproducibility (RSD) was 0618%.

Key words: HPLC, solasedine hydrochloride, identification, quality control