Development of Resonance Frequency Calibration Instrument of Overhauser Effect Sensor Electronic Paramagnetic

NIE Yahao, ZHANG Shuang, CHEN Shudong   

  1. College of Electronic Science and Engineering, Jilin University, Changchun 130012, China
  • Received:2025-01-17 Online:2026-04-14 Published:2026-04-14

摘要:

针对 Overhauser 效应传感器自由基溶液的电子顺磁共振频率特性问题, 设计了一种测量自由基电子顺磁共振频率的电子顺磁共振仪。 主机选用DDS(Direct Digital Synthesizer)芯片作为 30 ~ 70 MHz 连续可调的射频信号源, 并采用高带宽的鸟笼线圈作为传感器射频谐振腔。 实验表明, 在射频功率为 1. 5 W、 极化时间为 10 s的情况下, 共振仪输出信号信噪比为 38, 能成功标定未知自由基溶液的电子顺磁共振频率。

关键词:

Abstract:

The sensitivity of the Overhauser magnetometer is closely related to the polarization degree of the free radical solution in the sensor, and the polarization degree of the solution depends on the matching degree between the radio frequency excitation frequency and the electron paramagnetic resonance frequency in the free radical solution. To study the electron paramagnetic resonance frequency characteristics of the free radical solution in the Overhauser effect sensor, an electron paramagnetic resonance instrument for measuring the free radical electron paramagnetic resonance frequency is designed. The host of the electron paramagnetic resonance instrument uses DDS(Direct Digital Synthesizer) chip as a continuously adjustable radio frequency signal source of 30 MHz ~ 70 MHz, and adopts a high-bandwidth birdcage coil as the sensor radio frequency resonant cavity. The experiment results shows that under the condition of radio frequency power of 1. 5 W and polarization time of 10 s, the signal-to-noise ratio of the resonator output signal is 38, and the electron paramagnetic resonance frequency of the unknown free radical solution can be successfully calibrated.