Evaluation of left ventricular function in maintained hemodialysis patients using real-time three-dimensional echocardiography

 WANG Ling1, WANG Jing1, ZHENG Min1, ZHANG Chun-Mei1, LI Chan2   

  1. 1. Department of Ultrasonography,First Affiliated Hospital,Harbin Medical University,Harbin 150001,China|2. Department of Ultrasonography,No.211 Hospital of PLA,Harbin 150001,China
  • Received:2010-04-08 Online:2010-07-28 Published:2010-07-28

摘要:

目的:以磁共振显像(MRI)技术为对照标准,探讨实时三维超声心动图(RT-3DE)评价维持血液透析(MHD)患者的左心室容积(LVV)和射血分数(LVEF)的可行性与准确性。方法:MHD患者48例,采用RT-3DE采集其LVV和LVEF,同时应用二维超声心动图Simpson’s法M-Teichholz法进行检测,将3种方法的检测值与MRI所检测的数据进行比较。结果:①RT-3DE容积检测值和MRI测量的LVV有良好的相关性(r=0.90,P<0.001),配对t检验表明,两者检测值之间差异无显著性(P>0.05)。②在心腔变形时,M-Teichholz、Simpson’s法所测左室舒张末期容积(LVEDV)和左室收缩末期容积(LVESV)均高于或低于RT-3DE检测值,差异均有显著性(均P<0.05)。结论:RT-3DE可准确测量正常心腔及心腔变形情况下的LVV,在心腔变形情况下,RT-3DE较M-Teichholz、Simpson’s 法测得的LVV及LVEF数据准确性高。

关键词: 维持性血液透析;实时三维超声心动图;左室容积;核磁共振成像

Abstract:

Objective Using magnetic resonance imaging(MRI) technique as control standard,to explore the feasibility and accuracy of true real-time three-dimensional volume quantitative techniques (RT-3DE) in evaluation of  the left ventricular volume (LVV) and left ventricular ejection fraction (LVEF) of the maintained hemodialysis (MHD) patients.Methods 48  MHD patients were enrolled in this study.RT-3DE was used to detect the LVV and LVEF,while  Simpson’s method and M-Teichholz method were used at the same time.The results of the three methods were compared with the data measured by MRI.Results ①The data  measured by  RT-3DE method   showed a highly positive correlation  with  the   LVV detected by MRI  (r=0.90,P<0.001),but there was no significant  difference between the two methods (P>0.05).②When heart cavity deformed,the results of left ventricular end-diastolic volume(LVEDV) and left ventricular end-systolic volume (LVESV) measured by M-Teichholz method and Simpson’[KG-*3]s method were higher or lower than the data measured by RT-3DE method,and the differences were statistically significant (all P<0.05).Conclusion RT-3DE can accurately measure the volume of normal cardiac chambers and the left ventricular cavity deformation volume. The result of RT-3DE has higher accuracy compared with Simpson’s method and M-Teichholz method.

Key words: maintained hemodialysis, real-time three-dimensional echocardiography, left ventricular volume, magnetic resonance imaging