Influence of porcelain firing and different manufacturing techniques in fit of Co-Cr porcelain crowns

ZHAO Ying, XIE Yu, HE Xi, ZHANG Yan, WEI Wei, LI Jiang   

  1. Department of Prosthodontics, StomatologyHospital, Jilin University, Changchun 130021, China
  • Received:2017-11-28 Online:2018-05-28 Published:2018-05-31

Abstract: Objective: To explore the influence of different manufacturing techniques and different scanning objects in the marginal and internal fits of the Co-Cr porcelain crowns after procelain firing,and to provide the basis for judging the influencing factors of longevity of restorations. Methods: Five extracted maxillary premolar were selected.According to the prepared tooth,silicone rubber and stone die restoration models,which were fabricated by CAD/CAM and selective laser melting(SLM),the restorations were divided into control group(traditinal casting),tooth-SLM group,silicone rubber-SLM group,stone die-SLM group,tooth-CAD/CAM group,silicone rubber-CAD/CAM group, and stone die-CAD/CAM group.Then all the restorations were fired with the porcelain.The marginal and internal fits of the specinens were measured by micro-compated tomography(Micro-CT). Results: Compared with control group,the AMD of restorations in tooth-SLM and CAD/CAM groups were decresed(P<0.05),and the MG of restorations in silicone rubber-SLM group was decresed(P<0.05).Compared with control group,the CA and AW of restorations in silicone rubber-SLM group were decresed(P<0.05),the AW of restorations in tooth-CAD/CAM group was the least;compared with control group, the AOT and OA of restorations in stone die-SLM group and CAD/CAM group were decresed(P<0.05),and the AOT and OA of restorations in silicone rubber-CAD/CAM were the least.Compared with before porcelain firing,the marginal fit and internal fit of restorations after porcelain firing were incresed(P<0.05).There was positive corrlation between the MG of restorations and fabricated techniques and scanning objects(P<0.05).There was positive corrlation between the average of internal fit of restorations and fabricated techniques and scanning objects(P<0.05). Conclusion: The porcelain firing can affect the marginal and internal fits of Co-Cr porcelain crowns.The restorations fabricated by SLM had a better marginal fit and the restorations fabricated by CAD/CAM has a better internal fit.

Key words: porcelain firing, selective laser melting, CAD/CAM, Co-Cr porcelain crowns

CLC Number: 

  • R783