Evaluation on corrosion resistance of dental prosthetic materials
 Ni-Cr alloy,Co-Cr alloy and pure Ti in Cola

GE Yan1,HE Xi2,ZHANG Feng-lan2,LI Jiang2,ZHAO Chao-yue2,LI Xiao-ju2,CHEN Zhen-yu2,WANG Xu3   

  1. 1.Department of Oral Mucositis, Stomatology Hospital,Jilin University,Changchun 130021,China;2. Department of Prosthodontics,Stomatology Hospital,Jilin University,Changchun 130021,China;3. Department of Stomatology,Hospital of Traditional Chinese Medicine of Baicheng City,Baicheng 137000,China
  • Received:2012-09-19 Online:2013-03-28 Published:2013-05-28

Abstract: Abstract:Objective  To study the electrochemical corrosion of Ni-Cr alloy,Co-Cr alloy and pure titanium(Ti) in Cola,and to provide  basis for choice of clinical materials.Methods Ni-Cr alloy,Co-Cr alloy and pure Ti were divided into 3 groups.The classic three-electrode system was used to measure thecorrosion potential (Ecorr) and polarization curves,the polarization resistance (Rp),and corrosion current density (Icorr)  in three groups in artificial saliva and Cola,and the results were analyzed by statistics.Results  The values of Ecorr and Rp of Ni-Cr alloy,Co-Cr alloy and pure Ti in Cola were lower than those in artificial saliva(P<0.05),and the values of Icorr of them in Cola were higher than those in artificial saliva(P<0.05). In Cola,the values of Ecorr and Rp were Ni-Cr alloy<Co-Cr alloy

Key words: dental alloy, artificial saliva, Cola, electrochemical corrosion