吉林大学学报(工学版) ›› 2017, Vol. 47 ›› Issue (1): 301-307.doi: 10.13229/j.cnki.jdxbgxb201701044

• Orginal Article • Previous Articles     Next Articles

TOF-MS overlapped peak separation approach based on characteristic peak matching

BAO Ze-min1, LIU Guang-da1, LONG Tao2, QIU Chun-ling1, TIAN Di1, LIU Dun-yi2   

  1. 1.College of Instrumentation & Electrical Engineering, Jilin University,Changchun 130061,China;;
    2.Beijing SHRIMP Center, Institute of Geology, Chinese Academy of Geological Sciences, Beijing 100037, China
  • Received:2015-12-23 Online:2017-01-20 Published:2017-01-20

Abstract: In order to solve the mutual interference problem among homogeneous peaks in Time-of-flight and Mass Spectrometry (TOF-MS) quantitative analysis, a spectral peak separation method was presented. First, according to the technical characteristics that those spectral peaks with close mass number have close shape, the peak positions were determined by the correlation algorithm between Gaussian curve and the original mass spectrum. Then, the individual single peaks were selected by the principle of “high amplitudes kept and low amplitudes left”. The characteristic peak was obtained by stacking and normalization of the single peaks. The power spectral density function was constructed by the characteristic peak, and positions and amplitudes of the overlapped peak were determined in accordance with the minimum principle, thus achieving the overlapped peak separation. The result obtained by the above characteristic peak matching method was compared with the results of Gaussian matching method and Lorentzian matching method respectively. When dealing with symmetric peaks with δ= 0.2, these results are basically consistent. However, when dealing with asymmetric peaks with δ= 1.6, the effect of characteristic peak matching method is much better than the other two methods. Experimental results show that the presented method can be used to separate the 29Si and 28Si1H overlapped peak, the separation resolution is improved from 0.371 to 0.519, while the original shape information of the peeks is maintained.

Key words: information processing, time-of-flight mass spectrometry(TOF-MS), characteristic peak, peak separation

CLC Number: 

  • TH843
[1] Benninghoven A. Chemical analysis of inorganic and organic surfaces and thin films by static time-of-flight secondary ion mass spectrometry (TOF-SIMS)[J]. Angewandte Chemie International Edition in English,1994,33(10):1023-1043.
[2] Stephan T. TOF-SIMS in cosmochemistry[J]. Planetary and Space Science , 2001,49(9):859-906.
[3] Tellez H,Druce J,Hong J E,et al.Accurate and precise measurement of oxygen isotopic fraction and diffusion profiles by selective attenuation of secondary ions(SASI)[J]. Analytical Chemistry,2015,87(5):2907-2915.
[4] Yang C, He Z Y, Yu W C, Comparison of public peak detection algorithms for MALDI mass spectrometry data analysis[J]. BMC Bioinformatics,2009, 10:1-13.
[5] Malyarenko D I, Cooke W E, Bunai C L, et al. Automated assignment of ionization states in broad-mass matrix-assisted laser desorption/ionization spectra of protein mixtures[J]. Rapid Communications in Mass Spectrometry, 2010, 24(1):138-146.
[6] Zhang Z M,Tong X,Peng Y,et al. Multiscale peak detection in wavelet space[J]. Analyst,2015, 140(23) :7955-7964.
[7] Oleg N P, Oleksandr M B. The peak shape model for magnetic sector and time-of-flight mass spectrometers[J]. International Journal of Mass Spectrometry,2010,295(1-2):1-6.
[8] Tracy M B,Chen H,Weaver D M,et al. Precision enhancement of MALDI-TOF MS using high resolution peak detection and label-free alignment[J]. Proteomics, 2008, 8(8):1530-1538.
[9] Abel M L, Shimizu K,Holliman M, et al. Peak-fitting of high resolution TOF-SIMS spectra: a preliminary study[J]. Surface And Interface Analysis, 2009, 41(4): 265-268.
[10] Eric F S,Nestor R,Richard D S. High mass measurement accuracy determination for proteomics using multivariate regression fitting: application to electrospray ionization time-of-flight mass spectrometry[J]. Analytical Chemistry,2003,75(3):460-468.
[11] Martin K, Johan S, Anders B,et al. Improved method for peak picking in matrix-assisted laser desorption/ionization time-of-flight mass spectrometry[J]. Rapid Communication in Mass Spectrometry,2004,18(11):1208-1212.
[12] Coombes K R, Koomen J M ,Baggerly K A,et al. Understanding the characteristics of mass spectrometry data through the use of simulation[J]. Cancer Informatics, 2005,1(1): 41-52.
[13] 李宝强,李翠萍,黄启斌,等. 基于小波变换的便携式质谱重叠峰解析方法研究[J]. 质谱学报,2015,36(3):199-205.
Li Bao-qiang,Li Cui-ping,Huang Qi-bin,et al. Research of portable mass spectrometer overlapped peak resolution method based on wavelet transform[J]. Journal of Chinese Mass Spectrometry Society, 2015, 36(3): 199-205.
[14] Savage P S, Armytage R M G,Georg R B,et al. High temperature silicon isotope geochemistry[J]. Lithos, 2014,190-191: 500-519.
[15] Chmeleff J,Horn I,Steinhoefel G,et al. In situ determination of precise stable Si isotope ratios by UV-femtosecond laser ablation high-resolution multi-collector ICP-MS[J]. Chemical Geology,2008,249(1/2): 155-166.
[1] YING Huan,LIU Song-hua,TANG Bo-wen,HAN Li-fang,ZHOU Liang. Efficient deterministic replay technique based on adaptive release strategy [J]. Journal of Jilin University(Engineering and Technology Edition), 2018, 48(6): 1917-1924.
[2] LIU Zhong-min,WANG Yang,LI Zhan-ming,HU Wen-jin. Image segmentation algorithm based on SLIC and fast nearest neighbor region merging [J]. Journal of Jilin University(Engineering and Technology Edition), 2018, 48(6): 1931-1937.
[3] SHAN Ze-biao,LIU Xiao-song,SHI Hong-wei,WANG Chun-yang,SHI Yao-wu. DOA tracking algorithm using dynamic compressed sensing [J]. Journal of Jilin University(Engineering and Technology Edition), 2018, 48(6): 1938-1944.
[4] YAO Hai-yang, WANG Hai-yan, ZHANG Zhi-chen, SHEN Xiao-hong. Reverse-joint signal detection model with double Duffing oscillator [J]. 吉林大学学报(工学版), 2018, 48(4): 1282-1290.
[5] QUAN Wei, HAO Xiao-ming, SUN Ya-dong, BAI Bao-hua, WANG Yu-ting. Development of individual objective lens for head-mounted projective display based on optical system of actual human eye [J]. 吉林大学学报(工学版), 2018, 48(4): 1291-1297.
[6] CHEN Mian-shu, SU Yue, SANG Ai-jun, LI Pei-peng. Image classification methods based on space vector model [J]. 吉林大学学报(工学版), 2018, 48(3): 943-951.
[7] CHEN Tao, CUI Yue-han, GUO Li-min. Improved algorithm of multiple signal classification for single snapshot [J]. 吉林大学学报(工学版), 2018, 48(3): 952-956.
[8] MENG Guang-wei, LI Rong-jia, WANG Xin, ZHOU Li-ming, GU Shuai. Analysis of intensity factors of interface crack in piezoelectric bimaterials [J]. 吉林大学学报(工学版), 2018, 48(2): 500-506.
[9] LIN Jin-hua, WANG Yan-jie, SUN Hong-hai. Improved feature-adaptive subdivision for Catmull-Clark surface model [J]. 吉林大学学报(工学版), 2018, 48(2): 625-632.
[10] WANG Ke, LIU Fu, KANG Bing, HUO Tong-tong, ZHOU Qiu-zhan. Bionic hypocenter localization method inspired by sand scorpion in locating preys [J]. 吉林大学学报(工学版), 2018, 48(2): 633-639.
[11] YU Hua-nan, DU Yao, GUO Shu-xu. High-precision synchronous phasor measurement based on compressed sensing [J]. 吉林大学学报(工学版), 2018, 48(1): 312-318.
[12] WANG Fang-shi, WANG Jian, LI Bing, WANG Bo. Deep attribute learning based traffic sign detection [J]. 吉林大学学报(工学版), 2018, 48(1): 319-329.
[13] LIU Dong-liang, WANG Qiu-shuang. Instantaneous velocity extraction method on NGSLM data [J]. 吉林大学学报(工学版), 2018, 48(1): 330-335.
[14] TANG Kun, SHI Rong-hua. Detection of wireless sensor network failure area based on butterfly effect signal [J]. 吉林大学学报(工学版), 2017, 47(6): 1939-1948.
[15] LI Juan, MENG Ke-xin, LI Yue, LIU Hui-li. Seismic signal noise suppression based on similarity matched Wiener filtering [J]. 吉林大学学报(工学版), 2017, 47(6): 1964-1968.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
No Suggested Reading articles found!